Checking the levelling accuracy in the vertical position

For this check, you will need a free measuring distance on firm ground in front of a 10-metre tall wall.

  1. Fix a plumb line to the wall.
  2. Position the measuring tool in the vertical position on a firm, level surface.
  3. Switch the measuring tool on and allow it to level in.
  4. Set up the measuring tool so that the laser beam meets the plumb line at the exact centre of the upper end.
  5. The discrepancy d between the laser beam and the plumb line at the lower end of the line reveals the measuring tool's deviation from the vertical.

For a 10 m high measuring distance, the maximum permitted deviation is as follows: ±1 mm. The discrepancy d must therefore be no more than 1 mm.